Worked solution from the Cell structure MCQ booklet.
The question asks for the correct calibration calculation for each graticule unit.
The thick marks on the stage micrometer are 0.1 mm apart. Two gaps = 0.2 mm, and over this distance the graticule reads 54 units (about 27 units per gap).
One graticule unit = 0.2 mm ÷ 54. To change mm to µm, multiply by 1000.
So the calculation is 0.2 ÷ 54 × 1000, in µm.
So the answer is C.
A divides by 1000 and gives mm, which is wrong. B uses 52 units for one gap, which does not match the diagram. D uses 79 units.
▶ Whole paper
Video solution of the whole paper · by Prosperity Academy. Tap the picture to open it on YouTube.
Solutions follow the official Cambridge mark schemes. Videos are from other YouTube channels. Paper codes read component / series / year: M/J = May/June, O/N = October/November. Past-paper questions © UCLES, reproduced for classroom practice.